EasyManua.ls Logo

Siemens MAMMOMAT 1000 - Starting up the MAMMOMAT; Starting up the Workstation and Login; Starting up the MO Unit; Installation of Software

Siemens MAMMOMAT 1000
50 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Opdimar Register 3 SPB7-230.033.12 Page 2 of 4 Siemens-Elema AB
Installation Rev. 01 09.02 SPS-UD Solna, Sweden
6 - 2 Final procedures
Starting up the MAMMOMAT 6
Switch on the MAMMOMAT.
Starting up the workstation and login 6
The workstation is started according to instructions in Starting up the workstation and
login in Supplement to the Instructions for Use MAMMOMAT 3000 - Opdimar.
Use the user name “service” and the password obtained from Siemens service center
when entering the system.
To obtain optimal picture, adjust the monitor settings as follow:
Brightness: 50260
Contrast: 100
Starting up the MO unit 6
1. Insert the MO disk, type 2.3 GB MO disk
2. For switch settings, SCSI ID and termination see Ultra 10/Ultra 1 MO unit in
MAMMOMAT 1000/3000/3000 Nova - Opdimar Service Instructions.
Installation of software 6
The software is installed in the factory before delivery. If the software for any reason
needs to be reinstalled, refer to Reinstallation of software in MAMMOMAT 1000/3000/
3000 Nova - Opdimar Service Instructions.
Mounting of covers 6
Mount the covers previously removed.
Filling in of IVK list 6
Fill in the IVK list or report the IVK components to your responsible uptime service center
(USC).
HEWLETT
PACKARD
Fig. 1 Magneto-optical drive (HP)

Table of Contents

Other manuals for Siemens MAMMOMAT 1000