Opdimar Register 3 SPB7-230.033.12 Page 12 of 26 Siemens-Elema AB
Installation Rev. 01 09.02 SPS-UD Solna, Sweden
5 - 12 Installation
5. Connect the other end of the cable to the biopsy controller connector X206 in
Fig. 11.
6. Connect the MAMMOMAT printer (if applicable) to X205 on the biopsy controller
using the previously removed printer cable modified in the following way:
If the shield braid of the printer cable is bare where it was connected to the
generator bottom plate, the cable shall be insulated.
If connected, eject pin 1 on the printer connector side and insulate.
7. Strain relieve the cable at the bottom of the stand.
8. Connect the printer to the same wall socket (branch circuit) as the workstation.
The wall socket must be marked according to Fig. 24.
9. Close the cable duct cover and install the front cover of the generator.
The X205 connector is also used for connection of the service PC.
The switch S3 (TEST) on PC Board D702 is not used with Opdima.
The software will sense if a printer or a service PC is connected.
OPD00453
X201
X202
X203
X204
Ferrite
X205 X206 X207 Mains cable
Protective earth cable
Fig. 11 Connection of cables to the biopsy controller
NOTE
NOTE