16 Replacing basic unit components
MAMMOMAT Novation DR SPB7-250.841.01.04.02 Siemens AG
05.05 CS SD 24
Page 16 of 64
Medical Solutions
All files related to the old detector must first be archived, then the files related to the new
detector are uploaded. The procedures for the DROC and for the WH AWS are similar.
Actual steps may differ due to differences in the OS and application software.
For a DROC workstation on site:
1. Switch the MAMMOMAT Novation
DR
system ON.
First turn on the main power switch to initiate the BRICK startup.
After 5 minutes, turn on the control console.
After an additional 3 minutes, turn on the workstation.
2. There is no special procedure for turning the system off. Log in as root and enter
the appropriate password.
3. On the desktop, open File Manager and select the following directory:
/opt/linx_mp/opt/DRUL_0.6.3/data
It is recommended that the detector-specific files of the detectors used on each
MAMMOMAT Novation
DR
system be kept on the hard disk of the workstation. Before
uploading the detector-specific files of the new detector from the CD-ROM to the
workstation, the files for the removed detector must be stored. Use the following
procedure:
4. In the /opt/linx_mp/opt/DRUL_0.6.3/data directory, create a new subdirectory with
the name: MPwxyz , where wxyz are the four digits of the serial number of the re-
moved detector.
5. Locate the detector-specific files within this directory
/opt/linx_mp/opt/DRUL_0.6.3/data - they are:
MPwxyz.cal
MPwxyz.cfg
MPwxyz.map
MPwxyz.smj
6. Select (highlight) and cut these files from the directory
/opt/linx_mp/opt/DRUL_0.6.3/data and paste them into the newly created subdirecto-
ry MPwxyz.
7. Open a second File Manager.
8. Insert the supplied CD-ROM into the compact disk drive of the workstation.
9. Open the directory on the CD-ROM and locate the three detector-specific files for the
new detector:
MPabcd.cfg
MPabcd_final.map
MPabcd.smj
There should be three files. The .cal file is not present. It is first generated on the
MAMMOMAT Novation
DR
system during one of the following work steps, i.e., the
detector gain calibration.
10. Copy these three detector-specific files from the CD-ROM into the directory
/opt/linx_mp/opt/DRUL_0.6.3/data.