Siemens AG SPB7-250.841.01.04.02 MAMMOMAT Novation DR
05.05 CS SD 24
Replacing basic unit components 15
Page 15 of 64
Medical Solutions
7. Remove the leads from connector X 845 on the right side. The polarity is important. The
two leads are color coded; the red cable is on the top and the gray lead is down below.
NOTE
THE POLARITY OF THESE CONNECTORS (X846/X845) IS IMPOR-
TANT. MAKE SURE THAT THEY ARE CONNECTED CORRECTLY!
8. Remove the grid sliders on the right and on the left (1/Fig.5/p.14) by sliding them to-
wards the stand out of the Teflon rails.
9. Loosen and unscrew the two screws and remove the right transversal (lateral) grid slid-
er (1/Fig.6/p.15). This is necessary to be able to maneuver the main cable connector
out from under the right longitudinal rail. This allows the detector main cable to be re-
moved easily through the square recess (2/Fig.6/p.15) in the detector carrier plate.
10. Push the connector of the detector main cable out under the right longitudinal rail.
11. First, slightly loosen the four Allen screws that hold the detector in place
((3/Fig.4/p.14) and (3/Fig.6/p.15)). Prepare a place to put the detector after remov-
al; be careful not to drop the detector.
Fig. 6: Remove detector
12. Unscrew and remove the four Allen screws and remove the detector.
This concludes the disassembly and removal of the detector.
Installing the new detector
The mechanical installation of the new detector should be performed according to the
Installation Instructions or Startup manual, available in the technical documentation
folder delivered to the site.
For installation information, refer to:
whichever is applicable. The instructions for disassembly, properly applied, can also pro-
vide some guidance.
Updating the detector-specific files
• Chapter 4 - Detector installation of the System Startup with WH AWS, or to:
• Chapter 7 - Detector installation of the Installation Instructions and Start-Up
(DROC).