24 Replacing basic unit electrical assemblies
MAMMOMAT Novation DR SPB7-250.841.01.04.02 Siemens AG
05.05 CS SD 24
Page 24 of 64
Medical Solutions
2. To view the BRICK AEC configuration information, click the Brick AEC Configuration
link on the BRICK Control Panel page or the AEC link on any of the other PXCM pages.
The Brick AEC - Status page is displayed, as shown in (Fig.7/p.24).
Fig. 7: BRICK status
3. Note the corresponding effective values in (Tab. 1 / p. 24).
Tab. 1 BRICK values
Dismount the BRICK
• Turn OFF the entire system by switching OFF the mains input switch S700.
• Remove the service cover at the back of the MAMMOMAT.
• Remove all connectors and cables from the BRICK.
• Remove the BRICK. Be aware of the weight of the BRICK and exercise appropriate
caution.
Item Value
Brick AEC High Dose Scaling Factor
Brick AEC Low Dose Scaling Factor