54 Grid
MAMMOMAT Novation DR SPB7-250.841.01.04.02 Siemens AG
05.05 CS SD 24
Page 54 of 64
Medical Solutions
3. Adjust the grid switch so that it activates and deactivates on either side of the slits
((Fig. 30 / p. 54) and (Fig. 31 / p. 54)). You’ll hear a "click" when the switch activates or
deactivates.
Fig. 30: Adjusting the switch on edge 1
Fig. 31: Adjusting the switch on edge 2
4. Tighten the three screws.
5. Move the grid to the other end position and check that the grid switch activates/deacti-
vates on either edge of the slits ((Fig. 30 / p. 54) and (Fig. 31 / p. 54)).
Adjusting the grid location
1. Open the carbon detector cover.
2. Make sure the grid is mounted parallel to the detector and is lined up with the front of the
table.
The front switch and the springs for the grid holders (left and right) have to be engaged,
see System Start-up, SPB7-250.815.., the Detector Installation section for more
detailed information.