EasyManuals Logo

Siemens RAPIDLAB 248 User Manual

Siemens RAPIDLAB 248
135 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #22 background imageLoading...
Page #22 background image
Calibration Gases
Two gas standards are used to calibrate the pCO
2
and pO
2
sensors.
Cal Gas provides the calibration point for 1 and 2 point pCO
2
and
pO
2
calibrations. Cal Gas contains 5.00 ± 0.05% carbon
dioxide and 12.00 ± 0.05% oxygen, balanced with
nitrogen and is NBS traceable.
Slope Gas provides the slope point for 2 point pCO
2
and pO
2
calibrations. Slope Gas contains 10.00 ± 0.05% carbon
dioxide balanced with nitrogen and is NBS traceable.
WARNING Compressed gas cylinders require careful handling. To prevent
damage and possible personal injury, observe the following precautions:
Never drop cylinders, allow them to strike each other or subject
them to other strong shocks.
Secure cylinders to a wall or bench, or place in a cylinder base
support stand.
Never drag, roll or slide cylinders, even for short distances. Use a
suitable hand truck to move cylinders.
Never tamper with safety devices in regulators or cylinders.
Use these gases for the calibration of clinical and research
instrumentation only. US Law prohibits dispensing these gases for
drug use.
The contents are under pressure – do not puncture.
Do not use or store near heat or open flame.
Do not expose cylinders to temperatures above 54°C (130°F) as this
may cause the contents to vent or explode.
Never throw cylinders into fire or incinerators. Follow the disposal
instructions on the cylinders.
Calibration
The 248 automatically calibrates using one of the following user-
selectable methods. Five minutes before a calibration is due a
countdown message appears on the Ready screen. Samples can still be
measured during this time.
Fixed Time
The 248 automatically calibrates at a specified time interval. The
interval is user-selectable, and can be 30 or 60 minutes. A 1 point
calibration is carried out at every interval. A full 2 point calibration is
carried out at every fourth interval.
For example:
When the interval is set to 30 minutes, the 248 automatically carries
out a 1 point calibration every 30 minutes, and a 2 point calibration
every 2 hours.
Page 1-12 248 Operator’s Manual

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Siemens RAPIDLAB 248 and is the answer not in the manual?

Siemens RAPIDLAB 248 Specifications

General IconGeneral
BrandSiemens
ModelRAPIDLAB 248
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals