EasyManuals Logo

Siemens RAPIDLAB 248 User Manual

Siemens RAPIDLAB 248
135 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #55 background imageLoading...
Page #55 background image
Refilling or Replacing the Measurement Sensors
BIOHAZARD: See Appendix A, Protecting Yourself From Biohazards.
Equipment: pH fill solution (Cat. 478533) as required; replacement
sensors, as required; disinfectant (for example, Cat. 673390).
NOTE: Although the pCO
2
and pO
2
sensors contain fill solution they
cannot be refilled. To replace the pCO
2
and pO
2
sensors, follow the
instructions in steps 1 - 4 and 6 - 8.
1. Use the Disinfect routine and then stop the 248 system, Page 3-8.
2. Raise the front cover.
3. Slide the measurement block catch down and raise the block cover.
Figure 3-16. Opening the Measurement Block Cover
4. Swing and hold the tensioner to the right and remove the
appropriate sensor.
Figure 3-17. Removing a Sensor
5. Refilling the pH sensor:
CAUTION
Make sure you use pH fill solution. Do not use reference sensor fill
solution.
Do not touch the inner electrode as it is fragile and easily damaged.
a. Unscrew the inner electrode and set aside on lint-free tissue.
b. Empty the fill solution out of the sensor.
Maintaining the System Page 3-17

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Siemens RAPIDLAB 248 and is the answer not in the manual?

Siemens RAPIDLAB 248 Specifications

General IconGeneral
BrandSiemens
ModelRAPIDLAB 248
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals