EasyManuals Logo

Siemens RAPIDLAB 248 User Manual

Siemens RAPIDLAB 248
135 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #17 background imageLoading...
Page #17 background image
Understanding the System Page 1-7
Setup Options
You can configure the 248 by choosing units of measurement, setting
reference ranges and so on. This is covered in Section 5, Configuring the
System. Section 5 also gives recommended settings where applicable.
You can, of course, use the 248 using the default (factory set) options and
values. These are given below.
Operating Setup
QC
pH 6.001 - 8.000 (10.0 - 997.7 nmol/L H
+
)
pCO
2
5.0 - 250.0 mmHg (0.67 - 33.33 kPa)
pO
2
0.00 - 749.0 mmHg (0.00 - 99.86 kPa)
Reference Ranges
pH 7.350 - 7.450 (35.5 - 44.7 H
+
nmol/L)
pCO
2
35.0 - 45.0 mmHg (4.67 - 6.00 kPa)
pO
2
80.0 - 100.0 mmHg (10.67 - 13.33 kPa)
Units
pH pH units
Gases mmHg
ctHb g/dL
Calibration
method and interval flexible time, 30 minutes
gas values
cal 5% CO
2
12% O
2
slope 10% CO
2
0% O
2
Printer Options
printer on
print results only
Correlation
pH slope 1.000
pH intercept 0.000
pCO
2
slope 1.000
pCO
2
intercept 0.000
pO
2
slope 1.000
pO
2
intercept 0.000
CAUTION: These
values must not
be changed
without first
referring to
Section 5,
Configuring the
System.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Siemens RAPIDLAB 248 and is the answer not in the manual?

Siemens RAPIDLAB 248 Specifications

General IconGeneral
BrandSiemens
ModelRAPIDLAB 248
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals