Page 2-14 248 Operator’s Manual
Entering Standby Mode
This routine conserves reagents. During Standby the sensors are kept
wet and the pump tubes moved from time to time to keep them in good
condition. The 248 does not calibrate while in standby mode, but
automatically calibrates as required when restarted, before allowing
sample measurements.
1. From Ready press # for menu and 7 for Standby.
2. You can restart the 248 from standby mode in two ways:
a. press * to restart the 248 immediately, or
b. set an auto restart time. If a time is set, the 248 automatically
restarts at that time.
Setting Auto Restart Time
1. To set the auto restart time press #.
2. Enter the time required and press * to Exit. If the time displayed is
the time you want for auto restart, just press *.
Standby
Press * to restart now
press # to set auto restart time
11:03:48
press * to Exit
11:03:55
Standby
press # to cancel auto restart
11:04:10
Auto restart at 08:30
press * to restart now
Set Auto Restart Time
Auto restart at
08:30

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Siemens RAPIDLAB 248 and is the answer not in the manual?

Siemens RAPIDLAB 248 Specifications

General IconGeneral
BrandSiemens
ModelRAPIDLAB 248
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals