Configuring the System Page 5-5
Setting the Maintenance Prompts
The 248 displays prompts on the Ready screen to empty the waste
bottle and to deproteinize and condition the sensors. The prompts
appear at approximately 6.00 am at the intervals you select.
To set the maintenance prompts: from Ready press # for menu,
6 for System Setup and 2 for Maintenance Prompt.
The prompts can be set between 0 and 9 days. To cancel the prompts
press C to clear the value or enter 0.
Default setting: empty waste bottle every day
deproteinize/condition sensors every 7 days
Selecting Parameters
You can choose parameters for measurement and for calculation.
To select parameters: from Ready press # for menu, 6 for System
Setup and 3 for Parameters.
To select measured parameters press 1. Choose from:
all three channels selected
any two of the three channels selected
any one of the three channels selected.
selected
not selected
NOTE: The 248 will not allow you to turn off all three channels.
NOTES: Calculated parameters will only be displayed if all three
measurement channels are selected.
O
2
CT will only be displayed if ctHb is entered. pO
2
(A-a) and
pO
2
(a/A) will only be displayed if F
I
O
2
is entered.
To select calculated parameters press 2. Choose from:
HCO
3
actual
HCO
3
standard
BE(ecf) (formerly BE(vv)
BE(B) (formerly BE(vt)
•O
2
CT
•O
2
SAT
ctCO
2
pO
2
(A-a) (formerly A-aDO
2
)
pO
2
(a/A) (formerly a/A ratio).
selected
not selected
Default setting: all three measurement channels selected, no calculated
parameters selected.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Siemens RAPIDLAB 248 and is the answer not in the manual?

Siemens RAPIDLAB 248 Specifications

General IconGeneral
BrandSiemens
ModelRAPIDLAB 248
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals