Daily
• check calibrant levels and ‘change by’ date, shake buffer pack • wipe sample area, drip tray and external
surfaces • check drip tray is fitted and connected properly •
Weekly
• check fill solution in pH and reference sensors, and debubble sensors • deproteinize/condition sensors
• check gas cylinder pressure and flow rate •
Quarterly
• Use Disinfect routine • replace pump tubing, clean and lubricate rollers • replace drip tray if difficult to
clean •
Six-monthly
• replace bottle tubing •
Maintaining the System Page 3-3