EasyManua.ls Logo

Siemens RAPIDLAB 248 - Page 58

Siemens RAPIDLAB 248
135 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Replacing the Reference Sensor Cassette
Figure 3-21. Filling the Internal Reference Compartment
NOTE: Use the hex tool supplied to remove the reference electrode inner
and reservoir cap.
CAUTION: Do not touch the reference electrode inner as it is fragile and
easily damaged.
c. Remove the electrode inner from the old cassette, or, if you are
installing a new reference inner remove it from its container, and
screw into the new internal reference compartment. Do not
cross-thread the electrode inner.
Figure 3-22. Changing the Reference Inner
d. Inject the remaining solution into the reservoir up to the fill line,
and replace the reservoir cap until finger tight.
Figure 3-23. Filling the Reference Reservoir and Replacing the Cap
e. Tilt the reference cassette and tap the front with your knuckle to
remove air bubbles.
Page 3-20 248 Operator’s Manual
Old Cassette
New Cassette
New Cassette

Table of Contents

Related product manuals