Replacing the Pre-heater Tube
Figure 3-41. Removing the Pre-heater Tube
9. Ease the pre-heater tube under the plastic moulding.
10. Re-assemble using the new pre-heater tube assembly. Make sure that:
The pre-heater tube is in the pre-heater groove, and the block
connector is in place.
The reagent inlet connector is in place.
The pre-heater cover and sample detector cover are replaced.
The reagent manifold connector is connected.
The sensors are replaced and seated correctly, and the block cover is
closed.
11. Replace the probe and probe housing, Page 3-26, steps 12 - 16.
12. Lower the front cover and restart the 248.
Page 3-30 248 Operator’s Manual
Reagent Inlet
Connector
Block Connector
Pre-heater Tube
Plastic Moulding