EasyManuals Logo

Siemens RAPIDLAB 248 User Manual

Siemens RAPIDLAB 248
135 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #85 background imageLoading...
Page #85 background image
Measurement Block Routine
Cal Gas pCO
2
(mV) Slope Gas pCO
2
(mV)
Nominal mV –170.0 –151.0
Total pull in range -300.0 to +100.0 12.8 to 21.1 above
cal gas mV value
Action limits < -270.0 or >+80.0 <13.5 or >20.2 above
cal gas mV value
Cal Gas pO
2
(pA) Slope Gas pO
2
(pA)
Nominal pA +764.0 +80.0
Total pull in range 171 to 1961 above –100.0 to +250.0
slope gas pA value
Action limits <300 or >1600 above <–50 or >200
slope gas pA value
Action: pH - may be caused by the reference sensor, see Pages 4-1 and 4-2. If
problem persists condition or refill the pH sensor. If the problem is still
apparent, replace the pH sensor.
pCO
2
/pO
2
- deproteinize/replace sensor
* at 760 mmHg atmospheric pressure, using standard gases (Cal: 5% CO
2
,
12% O
2
and Slope: 10% CO
2
and 0% O
2
)
4. Stability – for 7.3 Buffer and Cal Gas a typical sensor will show the
following performance:
pH pCO
2
pO
2
Noise - after 15 seconds
the display will not
change by more than: 0.2 mV/10 secs 1.0 mV/10 secs 2 pA/10 secs
Drift - after 15 seconds
the display will not
change unidirectionally
by more than: 1.0 mV 5.0 mV 15 pA
during the remainder of the measurement.
Instability on the pH channel may be caused by the reference sensor.
De-bubble the reference sensor and repeat the test.
5. Press * to cancel the test.
Run Test Sample
1. You can use the run test sample option to:
a. Measure a sample with a known mV value (for example, run 7.3
or 6.8 buffer as a sample).
b. Run QC samples to promote stability after installing a new
sensor.
NOTE: You cannot use capillary samples with the Run Test Sample
routine.
Troubleshooting the System Page 4-11

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Siemens RAPIDLAB 248 and is the answer not in the manual?

Siemens RAPIDLAB 248 Specifications

General IconGeneral
BrandSiemens
ModelRAPIDLAB 248
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals