SC 7000 and SC 9000XL Patient Monitors Service Manual
ASK-T898-03-7600 Siemens Medical Systems, EM-PCS Danvers 131
7k9kXLSM.apc.CD_ROM.fm/04-99/kaupp
NOT A CONTROLLED DOCUMENT
80373xxx
80374xxx
MIB - related Verify proper SW version(s)
installed in all related
devices, in accordance with
SW compatibility chart for
Monitor SW.
If problem persists, try
disconnecting/reconnecting
cable(s) between Monitor
and MIB converter.
If problem persists,try
replacing cable(s) between
Monitor and MIB converter.
If problem persists, try
swapping MIB converter with
known-good device.
If problem persists, try
swapping CPS/IDS with
known-good device.
If problem persists, replace
Main Processor Board
Subassembly.
80375xxx
80376xxx
ISD (SDC) - related Verify proper SW version(s)
installed in all related
devices, in accordance with
SW compatibility chart for
Monitor SW.
If problem persists, try
disconnecting/reconnecting
cable(s) between Monitor
and ISD (SDC).
If problem persists,try
replacing cable(s) between
Monitor and ISD (SDC).
If problem persists, try
swapping ISD (SDC) with
known-good device.
If problem persists, try
swapping CPS/IDS with
known-good device.
If problem persists, replace
Main Processor Board
Subassembly.
Table C-1Diagnostic Log Codes (Continued)
Code
xxx = any
alphanumeric
characters
Description Cause Action