EasyManuals Logo

Siemens SC 7000 Service Manual

Siemens SC 7000
160 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #149 background imageLoading...
Page #149 background image
SC 7000 and SC 9000XL Patient Monitors Service Manual
ASK-T898-03-7600 Siemens Medical Systems, EM-PCS Danvers 133
7k9kXLSM.apc.CD_ROM.fm/04-99/kaupp
NOT A CONTROLLED DOCUMENT
8036f000 Bad Network Config
Data Reset
Upgrade SW to VC1.x.
8038d001 MIB Connection/
Disconnection reset
Upgrade SW to VC2.x.
803c1000 Remote view, auto
dual view reset
Upgrade SW to VC1.x.
803f1xxx
803f2xxx
803f3xxx
803f4xxx
etCO2 Pod/Cartridge - related Try swapping out airway
adaptor with known-good
components.
Try disconnecting/
reconnecting cable.
If problem persists, swap
cable with known-good
cable.
If problem persists try
swapping pod.
If problem persists, replace
Main Processor Board
Subassembly.
ffffffff boot_program_images failed with
status: A on image: B
Flash programming
error when
performing a
download
Retry download. If problem
persists replace Main
Processor Board
Subassembly if downloading
monitor SW. replace CPS or
IDS, If downloading CPS/
IDS software.
Table C-1Diagnostic Log Codes (Continued)
Code
xxx = any
alphanumeric
characters
Description Cause Action

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Siemens SC 7000 and is the answer not in the manual?

Siemens SC 7000 Specifications

General IconGeneral
BrandSiemens
ModelSC 7000
CategoryMeasuring Instruments
LanguageEnglish