Page 48 SITRANS Probe LU (HART) – OPERATING INSTRUCTIONS A5E32337695
Parameters
P346 Serial Number
Displays the serial number of the instrument. The numbers stored in Index 2, followed by
the numbers stored in Index 1, give you the complete serial number.
1. Select P346.
2. The primary reading displays one part of the serial number, with the secondary index
number visible in the auxiliary reading field.
3. Press DISPLAY twice to focus control on the secondary index.
4. Key in the other index number and press ENTER .
5. The other part of the serial number, associated with the new secondary index,
appears.
6. Press DISPLAY twice and use the ARROW keys or enter the value for the next
desired parameter.
Range Calibration (P650 to P654)
There are two possible types of calibration:
• Offset Adjusts the measurement by a fixed amount
• Sound Velocity Adjusts speed of sound, and changes the measurement
calculations
Do Offset calibration at any steady level, unless a Sound Velocity calibration is also done.
If both calibrations are done, then do Offset at a known high level, and Sound Velocity at
a known low level.
Index 2 Index 1
Values (view only)
Range: 00000 to 99999 Range: 00000 to 99999
Example: 1503010 15 03 010