A5E32337695 SITRANS Probe LU (HART) – OPERATING INSTRUCTIONS Page 53
Parameters
Enter a value slightly greater than the vessel’s maximum emptying rate, in Units (P005) or
percent of Span (P007) per minute.
Measurement Verification (P709 to P713)
P709 Damping Filter
Stabilizes the reported level within the Echo Lock Window (P713) in the event of level
fluctuations (for example, a rippling or splashing liquid surface). The value is in seconds,
and depends on the number of seconds it takes the device to reach 63% of a step value
change in reading.
The value is automatically altered when Measurement Response Speed (P003) changes.
The higher the value entered, the greater the range of stabilized fluctuation.
P711 Echo Lock
Use this feature to select the measurement verification process.
P003 Value Meters/Minute
10.100
2 1.0000
3 10.000
Values
Range 0 to 100.00 seconds (0 = off)
Default 10.000 seconds
Altered by
P003 Measurement Response
Related
P007 Span
P713 Echo Lock Window
Values
0
Off
1
Maximum Verification
2*
Material Agitator
3
Tot al Lock
Related
P700 Maximum Fill Rate
P701 Maximum Empty Rate
P712 Echo Lock Sampling
P713 Echo Lock Window
P820 Algorithm