4. Sanitization: Needs to be agreed with the customer.
Prior to initial startup, the water paths must be sanitized
with DENTOSEPT P as described in the instructions
for Care and Maintenance by the Practice Team.
If the customer needs the station for patients within
a week, you as the installation engineer have to ini-
tiate sanitization.
–If the customer does not need the station for
patients until later, the customer has to sanitize the
water system one day before using the station.
5. Technical documents
Keep the CUSTOMER'S COPY of the unit pass-
port together with the other technical documents.
The operating instructions supplied with the treat-
ment instruments should be kept in the plastic
jakket provided.
ATTENTION
The supplied technical documents are an important part of
the delivered unit and are thus to be handed out to the cus-
tomer. The technical documents must be kept by the cus-
tomer.