Loading...
63 22 726 D3561
D3561.201.01.03.02 05.2012
231
Sirona Dental Systems GmbH 10 Overview of all function keys
Operating Instructions SINIUS / SINIUS CS 10.4 Treatment program
båÖäáëÜ
10.4.2 Endodontics
Endodontics
Calibrating the burr drive
A calibration must be performed each time you change or lubricate the
contra-angle handpiece.
The contra-angle handpiece is automatically checked during calibration.
This includes a measurement of motor current at different speeds to
assess the properties of the system.
Direction of rotation
Switches counterclockwise rotation ON/OFF
Opening the sub-screen
Used to access additional subordinate functions; see the following
function keys:
ApexLocator and Auto-Reverse
Turning on/off the automatic motor stop of the ApexLocator
If your treatment center is equipped with the ApexLocator option, you can
determine that the burr drive automatically stops at the physiological
apex.
Auto reverse function
When the set torque value is reached, the burr drive automatically
switches to counterclockwise.
If your treatment center is equipped with the ApexLocator option, you can
determine that the burr drive automatically stops at the physiological
apex. If the Auto-Reverse function is switched on, the next time the foot
switch is activated following a motor stop, the motor is switched to
counterclockwise rotation. When the root canal file is withdrawn, the burr
drive automatically switches back to clockwise rotation.
Remove file from sequence
Deletes the selected files from the sequence.
Memory key
Stores the settings of an endodontic treatment
10.4.3 Endodontics administration
Endodontics administration
Opening the sub-screen
Access from the
Treatment selection
program. See the following function
keys:
Copy endodontic treatment
Copy treatment
A treatment can be copied and saved in the treatment list under a different
name. Then the settings can be changed.
Rename endodontic treatment
Endodontic treatments can be renamed to facilitate corrections and
editing.

Table of Contents

Other manuals for Sirona Sinius CS