Acoustic Output Parameter
D-2
Transducer Model:C3I60H Operating Mode:B
TIS TIB
Non-scan
Index Label
MI
Scan
A
aprt≤1cm² Aaprt>1cm²
Non-
scan
TIC
Global Maximum index value 0.845 0.278 0.635
Prα Mpa 1.523
P mW 42 42
Min.of[Pα(zs),Ita,α(zs)] mW
zs cm
zbp cm
zb cm
z at max. Ipi,α cm 4.05
deq(zb) cm
fawf MHz 3.244 3.244 3.244
X cm 1.92
Associated
Acoustic
parameters
Dim of
A
aprt
Y cm 1.12
td μsec 0.602
p
rr
Hz 8000
pr at max. Ipi Mpa 2.854
deq at max. Ipi cm
Other
information
I
pi,α at max.MI W/cm² 247.76
Control 1:
Power
dB 0 0 0
Control 2:
Focal Depth
cm 6.0 6.0 6.0
Control 3:
Scan Angel
。
70 70 70
Operating
Control
conditions
Control 4:
Frequency
MHz 3.5 3.5 3.5