Loading...
Acoustic Output Parameter
D-4
Transducer ModelC5I20H Operating ModeB
TIS TIB
Non-scan
Index Label
MI
Scan
A
aprt≤1cm² Aaprt>1cm²
Non-
scan
TIC
Global Maximum index value 0.700 0.177 0.743
Prα Mpa 1.507
P mW 38 38
Min.of[Pα(zs),Ita,α(zs)] mW
zs cm
zbp cm
zb cm
z at max. Ipi,α cm 2.65
deq(zb) cm
fawf MHz 4.635 4.635 4.635
X cm 1.20
Associated
Acoustic
parameters
Dim of
A
aprt
Y cm 1.07
td μsec 0.466
p
rr
Hz 8000
pr at max. Ipi Mpa 2.227
deq at max. Ipi cm
Other
information
I
pi,α at max.MI W/cm² 72.644
Control 1:
Power
dB 0 0 0
Control 2:
Focal Depth
cm 4.0 4.0 4.0
Control 3:
Scan Width
mm 110 110 110
Operating
Control
conditions
Control 4:
Frequency
MHz 5.0 5.0 5.0