4.10 INTERRUPTING A PROGRAM IN EXECUTION
INTERRUPTING A PROGRAM
A cycle in progress can be interrupted by pressing the Start/Stop button.
During the Suspension” of the program, the letter S” appears on the display, followed by the code of
the current program.
fig. 17 Interrupting a Program: hold down the Start/Stop button
For GW - Laboratory Glassware Washer series products, SUSPENSION phase:
1. If the temperature in the chamber is below 40°C: the door can be opened. If the door is closed again within 1
minute, the cycle can be restarted from the phase where it was interrupted. The cycle is always restarted by
holding down the Start/Stop button for 2 seconds.
2. If the temperature in the chamber is above 40°C: a reset cycle is required since the door cannot be opened.
3. If the cycle is interrupted and the user does not press any buttons within 1 minute: the appliance
automatically switches to Reset mode
fig. 18 Display screen, Suspension of program 5: S 05 “.