OMtAjJLer h
-
6.17
OUTPUT PARAMETER
SELECTION
Command
Parameter
Pari
Par 2
PXO
PYO
CE
Counter electrode.
PX1
PY1 RE1
Reference electrode
1.
PX2
PY2
RE2 Reference electrode
2.
PX3
PY3 ARE V
RE1'
V
RE2-
PX4
PY4
ARE-Bi
A^E'^BIAS"
PX5 PY5
I
Cell Current.
PX6 PY6
I-Bi Cell current
-
bias current.
PX7 PY7
5ARE/5I Polarization
resistance.*
PX8
PY8 5I/8ARE
Polarization
conductance.*
PX9 PY9
POL D.C.
polarization.
PX10
PY10
IPOL POL + D.C. bias.
PX11 PYU
DVM Zero
(for nulling).
*
Available
only with sweep.
I
!
|
6.18 HISTORY
FILE FUNCTIONS
Command
Function
FLO
Closes the file.
FL1 Opens the
file.
FS
/
Defines the file
size (number of results).
7=1
to
-450
(default=200)
UF
/
Defines the
location from which the
file is to be updated
and
outputs
record to
i/f.
VF1
Clears the file, i.e. erases the file
contents.
VF2
Outputs the
contents of the file to the interface(s).
?FP0
Returns the number
of results filed.
(0
to
450).
?FP1 Returns
the file pointer (location of last result
filed).
?NR
Returns the number
of readings taken.
(0
to
99999).
FDO
Controls the format
of the data output to the display
by the
’
VF2’
command. Outputs file to display
as
:
pari, par2.
FD1
as time, pari.
FD2
as time, par2.
CSB/S11287 User Guide/Issue AA4.22