1 Safety
Basic User Manual 3
The video printer should be connected to the specific interface by using the cable
provided by the manufacturer. Otherwise, electrical shock may occur.
Within the environment that is 1.8 meters (6 feet) around a patient, connect peripherals
to the auxiliary power outlet which is capable of isolation protection; or, power the
peripherals by the auxiliary output cable or the isolation transformer complied with
EN/IEC 60601-1, or the power input of the same safety level.
Within the patient environment, when removing the covers or the connectors from the
non-medical electrical equipment without tools during maintenance, calibration or
other operations, do not touch the parts of the non-medical electrical equipment and
the patient simultaneously.
Use only the probes provided by the manufacturer. Otherwise, the system will not
perform, and an accident such as a fire may result in the worst case.
Only the peripherals and accessories provided or recommended by the manufacturer
can be used. Using other devices or accessories may degrade the system performance
and even cause electrical shock.
Do not pour any fluid onto the system surfaces, as fluid seepage into the electrical
circuitry may cause excessive electrical current leakage or system failure. If any
water is spilled onto the system carelessly, stop using the system and contact the local
distributor immediately.
The AC power plug for the system is a three-prong grounded plug and should never
be adapted to any two-prong outlet or by using an adapter. Connect the AC power
plug of the multiple socket-outlet to a hospital-grade power outlet.
If the system is transported to the operating environment with a great temperature
change, leave it for approximately 4 hours before powering it on. Ensure that the
temperature and humidity inside and around the system are equivalent before an
operation.
Do not use the system around a strong electric field, a strong electromagnetic field,
or the devices which generate radio waves, such as a radio, cellular telephones, or
transceivers. Using the system in an improper environment may result in malfunction
or damage.
Select the conformed multiple socket-outlet with protective grounding
,
and ensure
that its maximum output power exceeds the requirement of the system.
The multiple socket-outlet can only be used to provide power to the recommended
peripherals of the system.
Do not connect other devices to the multiple socket-outlet. Otherwise, the rated output
power of the multiple socket-outlet may be exceeded, and it may result in failure.