10 Probes and Biopsy Brackets
138 Basic User Manual
1. Clean the biopsy bracket.
For details, refer to Section 10.2.6 Cleaning the Biopsy Bracket.
2. Sterilize the biopsy bracket as shown in Table 10-5 for the entire recommended
soaking period.
3. Rinse the bracket with running sterile water to completely remove the residual
sterilant.
4. Dry the bracket with a lint-free soft dry cloth.
Table 10-5 Recommended Method for Sterilization
Sterilant Manufacturer Active
Ingredients
Level for
Active
Ingredient
Contact
Type
Contact
Period
Cidex™
Activated
Dialdehyde
Solution
J&J Glutaraldehyde 2.4% Soak 10 hours
10.2.8 Storage
Ensure the biopsy bracket is cleaned, sterilized and completely dried before storing it.
Store the biopsy bracket in a sterile environment and keep it away from the environment of
sudden temperature changes or direct sunlight.