138 Temperature and humidity limits
•DVDrecorder
Non-medical grade
•KensingtonSecurityCable
•USBkeyboard
Temperature and humidity
limits
Operating
System, battery, and transducer
10–40°C(50–104°F),15–95%R.H.
700to1060hPa(0.7to1.05ATM)
Shipping and storage
System and transducer
‐35–65°C(‐31 –149°F),15–95%R.H.
500to1060hPa(0.5to1.05ATM)
Battery
‐20–60°C(‐4–140°F),15–95%R.H.(Forstorage
longerthan30days,storeatorbelowroom
temperature.)
500–1060hPa(0.5–1.05ATM)
Electrical
Power Supply Input 100–240VAC,50/60Hz,
2.0–1.0A
Power Supply Output #1 15VDC,5.0A
Power Supply Output #2 12VDC,2.3A
Combinedoutputnotexceeding75watts.
Battery
Thebatterycomprisessixlithium‐ioncellsplus
electronics,atemperaturesensor,andbattery
contacts.
Runtimeisuptotwo hours,dependingon
imagingmodeanddisplaybrightness.
Electromechanical safety
standards
EN60601‐1:1997,EuropeanNorm,Medical
ElectricalEquipment–Part 1. General
RequirementsforSafety.
EN60601‐1‐1:2001,EuropeanNorm,Medical
ElectricalEquipment–Part1.General
RequirementsforSafety–Section1‐1.Collateral
Standard.SafetyRequirementsforMedical
ElectricalSystems.
EN60601‐2‐37:2001+AmendmentA1:2004+
AmendmentA2:2005,EuropeanNorm,
Particularrequirementsforthe
safetyof
ultrasonicmedicaldiagnosticandmonitoring
equipment.
CAN/CSAC22.2,No.601.1‐M90,Canadian
StandardsAssociation,MedicalElectrical
Equipment–Part1.GeneralRequirementsfor
Safety(includingCSA601.1Supplement1:1994
andCSA601.1Amendment2:1998).
CEI/IEC61157:1992,Interna tional
ElectrotechnicalCommission,Requirementsfor
theDeclarationoftheAcousticOutputof
MedicalDiagnosticUltrasonic
Equipment.
UL60601‐1(1stEdition),Underwriters
Laboratories,MedicalElectrical
Equipment‐Part 1:GeneralRequirementsfor
Safety.