8 Preparing transducers
•TheDELETEkeydeletesthecharacterright
ofthepointer.
3 (Optional)Informs,navigateamongtext
fields:
•ClickNexttoadvancetothenextfield.
•ClickPrevtoreturntothepreviousfield.
4 Toexitthe keyboard,clickoneofthe
following:
OKtosavechanges.
2Dtosavechangesanddisplay2D
imaging.
Preparing transducers
Acousticcouplinggelmustbeusedduring
exams.Althoughmostgelsprovidesuitable
acousticcoupling,somegelsareincompatible
withsometransducermaterials.SonoSite
recommendsAquasonic
®
gelandprovidesa
samplewiththesystem.
Forgeneraluse,applyaliberalamountofgel
betweenthetransducerandthebody.For
invasiveorsurgicaluse,applyatransducer
sheath.
ForinformationaboutpreparingtheTEEx
transducer,seetheTEExTransducerUserGuide.
To apply a transducer sheath
SonoSiterecommendstheuseofmarketcleared,
transducersheathsforintracavitaryorsurgical
applications.Tolessentheriskofcontamination,
installthesheathonlywhenyouarereadyto
performtheprocedure.
1 Placegelinsidethesheath.
2 Insertthetransducerintothesheath.
3 Pullthesheathoverthetransducerandcable
until
thesheathisfullyextended.
4 Securethesheathusingthebandssupplied
withthesheath.
5 Checkforandeliminatebubblesbetweenthe
faceofthetransducerandthesheath.
Bubblesbetweenthefaceofthetransducer
andthesheathmayaffecttheultrasound
image.
WARNING:
Some transducer sheaths contain
natural rubber latex and talc, which
can cause allergic reactions in
some individuals. Refer to
21 CFR 801.437, User labeling for
devices that contain natural rubber.
Some gels and sterilants can cause
an allergic reaction on some
individuals.
Caution:
To avoid damage to the transducer,
use only gels recommended by
SonoSite. Using gels other than the
one recommended by SonoSite can
damage the transducer and void
the warranty. If you have questions
about gel compatibility, contact
SonoSite or your local
representative.
SonoSite recommends that you
clean transducers after each use.
See Cleaning and disinfecting
transducers” on page 58.
WARNING:
To prevent contamination, the use
of sterile transducer sheaths and
sterile coupling gel is
recommended for clinical
applications of an invasive or
surgical nature. Do not apply the
transducer sheath and gel until you
are ready to perform the
procedure.