Chapter 4: Measurements and Calculations 47
Measurements
4 Usingthetouchpad,positionthesecond
verticalcaliperatthepeakofthenext
heartbeat.
5 Savethecalculation.(See“Tosavea
calculation”onpage 38.)
To calculate Cardiac Output (CO) or Cardiac
Index (CI)
TheCOandCIcalculationsrequireStroke
Volumeand He artRatecalculations.CIalso
requiresBodySurfaceArea(BSA).Afterthe
measurementsaresaved,theresultappearsin
thepatientreport.
1 (CIOnly)FillintheHeightandWeightfields
onthepatientinformationform.TheBSAis
calculatedautomatically.(See
“Tocreatea
newpatientinformationform”onpage 27.)
2 CalculateSV.See“TocalculateStrokeVolume
(SV)orStrokeIndex(SI)”onpage 46.
3 CalculateHR.See“TocalculateH eartRate
(HR)”onpage 46.
To calculate Cardiac Output automatically
Thesystemcanmaintaintheaccuracyof
automaticCardiacOutputmeasurementsonlyif
theflowrateis1L/minorgreater.
1 MeasurefromLVOT(2D):
a Onafrozen2Dimage,press Calcs.
b Fromthecalculationsmenu,selectCO,and
thenselectLVOT D.
c Positionthecalipers.(See“Workingwith
calipers”on
page 33.)
d Savethecalculation.(See“Tosavea
calculation”onpage 38.)
2 Traceautomatically(Doppler):
Theautomatictracetoolalwaysmeasuresthe
peakregardlessoftheLiveTracesettingin
Presetssetup.
a DisplaytheDopplerspectraltrace
(waveform).
b ForSweep Speed,selectSlowor
Med.
c Press Trace,andthenpressAboveor
Belowforthepositionoftheautomatic
tracetoolrelativetothebaseline.
Theautomatictrace toolappearsinyellow.
Theresultsappearatthebottomofthe
screen.
d Freezetheimage.
Ifyouwanttochangethewaveform
measured,moveeachverticalcaliperby
clickingandthenusingthetouchpad.
PressSettoupdatetheresults.
Ifyouinvertthefrozenimage,turnthe
cineknob ,ormovethebaseline,
resultsarecleared.
e Savethecalculation.
WARNING:
To avoid incorrect calculation
results, make sure that the Doppler
signal does not alias.
To avoid an incorrect diagnosis:
Do not use automatic Cardiac
Output calculations as the sole
diagnostic criteria. Use them only
in conjunction with other clinical
information and patient history.
Do not use automatic Cardiac
Output calculations in neonatal
patients.
To avoid inaccurate velocity
measurements if you use PW
Doppler, make sure that Angle
Correction is set to zero.