Chapter 6: Safety 71
Safety
• Before using the transducer, inspect the transducer face, housing, and cable. Do not
use the transducer if the transducer or cable is damaged.
• Always disconnect the power supply from the system before cleaning the system.
• Do not use any transducer that has been immersed beyond the specified cleaning
or disinfection level. See Chapter 5, “Troubleshooting and Maintenance.”
• Use only accessories and peripherals recommended by SonoSite, including the
power supply. Connection of accessories and peripherals not recommended by
SonoSite could result in electrical shock. Contact SonoSite or your local
representative for a list of accessories and peripherals available from or recommend
by SonoSite.
To avoid the risk of electrical shock and fire hazard:
• Inspect the power supply, AC power cords, cables, and plugs on a regular basis.
Ensure that they are not damaged.
• The power cord set that connects the power supply of the ultrasound system,
S Series stand, or V-Universal stand to mains power must only be used with the
power supply or stand, and cannot be used to connect other devices to mains
power.
To prevent injury to the operator/bystander, the transducer must be removed from
patient contact before the application of a high-voltage defibrillation pulse.
To avoid possible electrical shock or electromagnetic interference, verify proper
operation and compliance with relevant safety standards for all equipment before
clinical use. Connecting additional equipment to the ultrasound system constitutes
configuring a medical system. SonoSite recommends verifying that the system, all
combinations of equipment, and accessories connected to the ultrasound system
comply with JACHO installation requirements and/or safety standards such as
AAMI-ES1, NFPA 99 OR IEC Standard 60601-1-1 and electromagnetic compatibility
standard IEC 60601-1-2 (Electromagnetic compatibility), and are certified according
to IEC Standard 60950 (Information Technology Equipment (ITE)).
Caution:
Do not use the system if an error message appears on the image display: note the
error code; call SonoSite or your local representative; turn off the system by pressing
and holding the power key until the system powers down.
To avoid increasing the system and transducer connector temperature, do not block
the airflow to the ventilation holes on the back of the system.