Manual Integrator Supercal 531 06-06-2011 60
11. Technical Data
11.1 Integrator Supercal 531
CE-Conformity:
EMC-directive 89/336/EEC
Electro-magnetic immunity acc. to EN 50082-1
Electro-magnetic emission acc. to EN 50081-1
Low voltage directive 73/23/EEC for power supply units protection class *)
Mains 230 VAC II acc. to EN 61558
Mains 24 VAC III acc. to EN 61558
Battery III acc. to EN 61558
*) The protection class refers to the complete unit (integrator and connected power supply module)
Housing protection class: IP 65 acc. to EN 60529
Area class: A acc. to EN1434
Admissible ambient temperatures:
In operation + 5...55 °C
Transport and storage – 20...70 °C (dry)
Admissible ambient humidity < 93% relative humidity
Display: 8 digits
Energy 99’999’999 (kWh, MWh, MJ, BTU or GJ)
Volume 99’999’999 (m
3
, US-gallons)
Data storage non-volatile due to EEPROM
Temperature measurement:
Measuring element: Pt500 and Pt100 acc. to EN 60751
Temperature ranges:
Possible temperature range Θmin = -20 °C, Θmax = +200 °C
Admissible temperature range Θ2 °C...200 °C
Temperature difference ∆ Θmin = 1 K, ∆ Θmax = 150 K
Admissible temperature difference ∆ Θ 2...150 K
Response limit ∆ Θ 0.2 K
Temperature resolution 0.005 K
Measurement precision more precise than required by EN1434-1
11.2 Mains supplies
The Supercal 531 can be supplied by battery or mains modules. They can be retrofitted at any
time.
The mains modules are provided with a jumper. The backup-battery can be activated or deacti-
vated by this jumper. When commissioning the integrator 531, the jumper is plugged-in (activated).