Manual Integrator Supercal 531 06-06-2011 77
Overview of the test sequence integrator (simulated flow) ‚RKS / RTS’:
Entry of the number of measurements of simulation pulses (1 temperature measurement per s)
in the meter
NowaStart
NowaStop
Readout of measuring data, interim storage and transmission to test rig
Errors received by test rig
No calibration possible.
Overview of the test sequence temperature measurement ‚TK / TT’:
NowaStart
Instantaneous values received by test rig
NowaStop
Readout of measurement data, calculation of temperature, interim storage and transmission to
test rig
Errors received by test rig
No calibration possible.
16.1.1.3 Programme options and call parameters
Parameters compatible with test programmes from other manufacturers:
'/DEBUG' => extended debug displays on the screen
The programme options can be selected in three ways:
as entry in the file c:\SON\SC531\OPTIONS.CFG
as entry in the file c:\SON\OPTIONS.CFG
as call parameter in the command line
In the same sequence and priority, the programme searches for applicable options. If an option is
set several times (and possibly differently), the last stated value applies (e.g. of the command line).
The following options are possible:
NoMirror: Operation with optical head without echo
NoMux: No Nowa-multiplexer hardware connected, operation directly
over Com1 („standalone“)
Debug: NowaCom-protocol window active
Detection time = value [s]: During the set time the programme searches for meters (in case
they have been connected only after start of the programme)
Cmd = command file: Instead of the NowaCom-driver, the commands are read out
from the stated file („standalone“)
Prm = parameterisation file: In connection with a command file, the transfer of a file name is
admitted for the parameterisation without being stated explicitly
in the command file („standalone“)
RegDump: After detection of the meter, all registers stated in the register
file are read out and recorded in the file ‚REGSPlatznum-
mer.DMP’ (e.g. REGS1.DMP) (may take some time)
XP: Standalone-mode in WinXP (with the respective drivers)
In the file OPTIONS.CFG, the options can be registered line by line or within one line (with space
character or put in commas). Format of the command line for options: NOWA
[/SON:Option1,Option2,Option3,…etc.