CRANEX
®
Novus e 5. S2terminal and service commands
Service manual 85
NOTE:
The kV value used for the exposure will be the one
selected before entering the service mode. The kV
value cannot be changed while the unit is in the service
mode. If you wish to change the kV, you must exit the
service mode.
NOTE:
The "still" command, described later, generates x-rays
and activates the CCD sensor but the rotating unit
does not move.
expcnt exposure counter, use [<clear>] option to clear it
Displays the total number of exposures taken with the
tubehead.
To zero the exposure counter enter clear,
____________________________________________
>expcnt clear
____________________________________________
and then press Enter.
NOTE:
Always zero the exposure counter when the tubehead
is replaced.
help display info on command (‘help [cmd]’ or ‘h [cmd]’)
Displays the service command list and information
about the commands.
imagecalib takes a calibration image for calibrating the DSD
Creates a technical image which is then used by the im-
age capturing driver for image correction.
This function can also be carried out using the unit GUI.
See the section "Calibrating the CCD sensor" in the
installation manual.