86 Service manual
5. S2terminal and service commands CRANEX
®
Novus e
ip con gure ip address
Displays the current IP address of the unit.
To change the IP address enter the new one,
____________________________________________
> ip aaa.bbb.ccc.ddd (the new IP address)
____________________________________________
and then press Enter.
After changing the IP address of the unit, communica-
tion between the unit and the PC will be lost and the link
to the PC will have to be recon gured. Refer to section,
3.5 Con guring the communication link to the PC, in
the installation manual.
NOTE:
The IP address can also be reset using the driver tools
kVcheck test kV/mA combinations
Displays feedback values of all kV/mA combinations.
log service log [<all> <clear> <newest> <statistics>]
Displays exposure logs.
There are several log options:
<all> = (default) display the complete log
<clear> erase all log entries
<newest> display the latest log entry
<statistics> display the log statistics.
To select one of the log options enter the required com-
mand,
____________________________________________
> log clear
____________________________________________
and then press Enter.