CRANEX
®
Novus e 5. S2terminal and service commands
Service manual 87
logsign make a service signature for current events
Allows a short note to be added to a log entry.
Add the note,
____________________________________________
> logsign Note ......
____________________________________________
and then press Enter. The note will appear in the log.
mac print mac address
Displays the current MAC address of the unit.
mainscalib calibrates mains voltage
parameters: <current voltage>
Recalibrates the mains voltage if L1200 CPU board is
replaced.
Use a RMS multimeter to measure the actual supply
voltage (from the mains plug, L1910 or L1500) and then
enter this value.
____________________________________________
>mainscalib xxx
(xxx is the value read from multimeter)
____________________________________________
and then press Enter.
mainsprint print current mains voltage value
Displays the current line voltage value as measured by
the unit.
Useful for debugging when the unit displays error E5.
motortest run exposure ramp
A test drive for the rotation motor. The test checks that
motor is driven to the correct angle and that it works as
speci ed.