90 Service manual
5. S2terminal and service commands CRANEX
®
Novus e
stillmode still image exposure
Generates x-rays and activates the CCD sensor but the
rotating unit does not move.
It is used to check the beam alignment during unit
installation (refer to the Installation manual for more
information).
CAUTION:
Protect yourself from radiation when carrying out
this procedure.
Press and hold the exposure switch to take a still image
exposure.
NOTE:
The "exp" command, described earlier, generates x-
rays but does not activate the sensor or move the rotat-
ing unit.
warmup tubehead warmup before rst use
Initializes the tubehead. Must be used when the unit is
installed or if the unit has not been used for a long time.
CAUTION:
Protect yourself from radiation when carrying out
this procedure.
Press and hold the exposure switch until the procedure
is complete. The procedure may take some minutes.