CRANEX
®
Novus e 5. S2terminal and service commands
Service manual 89
printgen prints the generator settings
Displays the settings that are stored in the confi guration
memory.
quit quit service terminal
Exits and closes the service terminal.
reset reboot the unit
Reboots the unit.
restore restore factory confi guration
Restores the confi guration settings to the default factory
settings. You must key in restore confi guration.
____________________________________________
> restore confi guration
____________________________________________
and then press Enter.
setserno confi gure serial number (and MAC accordingly)
To display the CURRENT setting key in
____________________________________________
> serno => the current serial number appears
____________________________________________
and then press Enter.
To change the serial number, e.g. if you change the
CPU board key in
____________________________________________
> setserno SOXXXXXXX (new serial number)
____________________________________________
and then press Enter.