www.spacelabshealthcare.com
12-4
XPREZZON
AND QUBE
O
PERATIONS MANUAL
M ONITOR SETUP
Audio Off Reminder
By default, the control to turn off audio is not available to the clinician. Your system
administrator can select whether clinicians have access to turn off the audio. If
configured for access, you can turn on or off audio in the Audio tab.
If you have access to turning the audio on or off, the system administrator also has
the option to turn on or off an audio tone reminder. If enabled, a reminder tone
sounds every 60 seconds and a reminder message is displayed for 20 seconds.
There are two types of reminders (refer to Figure 12-2 and
Figure 12-3). If the SILENCE button is present, then you can silence the reminder
tone and close the reminder window by touching the button.
Figure 12-2 “Audio Off” Reminder Message
Caution:
If you touch the SILENCE button, the reminder tone and message
will deactivate until the monitor is powered off/on or until the
clinician turns the high priority alarm tone on again.
Figure 12-3 “Audio Off” Reminder dialog box

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Spacelabs XPREZZON and is the answer not in the manual?

Spacelabs XPREZZON Specifications

General IconGeneral
BrandSpacelabs
ModelXPREZZON
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals