PXS-710D Section 1
10
1.4 CONTINUED:
The unit has been designed to be used and installed on a flat surface as a
fixed unit that stays in place with no wheels.
It is the responsibility of the operator to always watch all parts of the unit to
verify that there is neither interference nor possibility of collision with the
patient or with other equipment.
Responsible organizations (e.g. X-ray System Integrators) combining
imaging components or accessories with the PXS-710D are required to insure
the resulting system meets the requirements of ANSI/AAME ES 60601-1 and
the standards identified in section 1.7.
Maximum Permissible Dose (MPD): Make sure that the X-ray generator
is set in working position with the reference axis (X-ray beam) pointing to the
reception area. The operator shall use the largest possible focal spot to skin
distance in order to keep the absorbed dose as low as reasonably achievable.
Trained personnel operating this unit should be familiar with the
Recommendations of the International Commission on Radiological
Protection, contained in Annals Number 60 of the ICRP with the applicable
National Standards and should have been trained in use of the equipment.
1.5
ELECTROMAGNETIC COMPATIBILITY WARNINGS & DECLARATIONS
1.5.1 INSTRUCTIONS FOR USE
[ENVIRONMENTS]
The PXS-710D is intended for use in a Professional Health Care Facility.
The PXS-710D may cause radio interference or may disrupt the
operation of nearby equipment. It may be necessary to take mitigation
measures, such as additional shielding or relocating or re-orienting the
equipment.