PXS-710D Section 5
66
UNIT MAINTENANCE
CAUTION
This section pertains to authorized Source-Ray service personnel only
Equipment installation and servicing procedures should be performed by
properly trained and qualified service personal only.
Protective covers are provided to shield service personal from inadvertent
contact with high voltage components.
Electrical shock hazard! The HF Series X-ray Generator contains the
serviceable components. Do not attempt to disable these components or
remove any trim covers. Refer service to qualified service personnel.
5.1 INTRODUCTION
In order to keep this Source-Ray unit operating properly, it must be inspected
and checked in accordance with the recommended intervals depending on the
age and use of the equipment. If there is a problem, call an authorized
serviceman.
DO NOT OPERATE THE EQUIPMENT UNTIL REPAIRS ARE COMPLETED.
5.2 RECOMMENDED INTERVALS
Periodic preventive maintenance including cleaning, mechanical and electrical
adjustments, are to be performed every six months or 3000 exposures.
Calibration intervals should not exceed 12 months. The periodic maintenance
schedule is required to maintain the unit in proper and accurate working
condition. Service life of unit is 5 years.