PXS-710D Section 5
67
5.3 ORDERING INFORMATION
To order replacement parts contact your local authorized Source-Ray dealer.
Authorized Source-Ray dealers should contact the main office at (631) 244-8200
between the hours of 8:00 AM and 5:00 PM. When ordering replacement parts
be
prepared to submit the following information: purchase order no. (If possible),
Source-Ray part number, part description, quantity required and shipping
instructions. If replacement parts are for a unit under warranty be prepared to
submit model numbers and serial numbers to expedite your order.
5.4 MAINTENANCE PROCEDURE
Reference the checklist on following page.
The PXS-710D contains no user serviceable parts.
5.5 WARM-UP PROCEDURE
If equipment has remained un-powered for an extended period of time, it will be
necessary to bring the unit up slowly so that the high voltage circuits are not
damaged.
1) No patient should be present during Warm-Up procedure.
2) Completely close shutters on the collimator.
3) During X-ray exposure operator should be located in a safe distance outside
the x-ray field, governed by the regulations and requirement of every state.
4) Turn the unit on and set the following; 60 kVp, Low mA and 10 mAs.
5) Take a 2 – 3 exposures.
6) Increase kV by ten and repeat step 3.
7) Repeat steps 2 and three until you reach the 110kV setting.