EasyManua.ls Logo

Sri PXS-710D - Settings Confirmation

Sri PXS-710D
73 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
PXS-710D Section 2
39
2.5.4
C
ONTINUED:
j. Make an Exposure and verify the actual kVp value, Repeat step i if
more adjustment is required.
k. Using the attached mAs meter, measure the mAs for the 1 sec.
exposure. Verify that the reading is 10.0 mAs (1.0 mAs +/- 5%). If
adjustment is required, adjust R102 on the Control board assembly
P/N 950132. Inspect squareness of wave-shape and repeat steps f -
g above if necessary.
l. When the calibration is completed for the 40 kVp preheat range,
depress the “NEXT” button to move to the next calibration station.
Repeat steps f – k. This process is repeated for 50, 60, and 70 kVp,
as displayed in the kV window.
m. Once the 70 kVp station is calibrated, pressing “NEXT” results in a
confirmation dialog as in fig 2.7. Click on the ‘OK’ button to proceed.
F
igure 2.7
S
ettings
C
onfirmation

Table of Contents