PXS-710D Section 5
68
TABLE 5.1
MAINTENANCE CHECKLIST (EVERY SIX MONTHS)
MODEL: PXS-710D SERIAL # ___________
1) Beam Current (Reference Section 2.3) ____________
2) Tube Potential (Reference Section 2.4). ____________
3) Exposure time (Reference Section 2.5). ____________
4) Beam Alignment (Reference Section 2.6) ____________
5) Check voltage on X-Ray Generator at J2 pins A and B
(collimator supply) with collimator unplugged (24 VDC). ____________
6) Check that all Indicator lights are operating. ____________
7) Check that all hardware is tightly secured, tighten if required. ____________
8) Check that collimator-mounting ring is tight. ____________
9) Check all interconnect cables for broken or frayed wires. ____________
10) Check that all hardware is secure and properly fastened. ____________
11) Check that all friction locks (head rotation) are properly
adjusted, tighten as required. ____________
12) Inspect spring mounts. Make sure all mounting
hardware is secure. (tighten if required) ____________
13) Inspect spring strength, if unit drifts down when lock is not
engaged, replace piston. ____________
NOTES:
When cleaning, use a cloth moistened in warm soapy water (use mild soap). Do
not soak the unit and do not use strong cleaners or solvents as they may blur the
lettering or dull the finish. Polish with pure liquid or paste wax.
Do not use a wax containing a cleaning substance. Do not polish the meter
windows.
SERVICE ENGINEER: ____________________ DATE: _____________