PXS-710D Section 5
69
TABLE 5.2
DISPLAYED FAULT CONDITIONS
DISPLAY FAULT DESCRIPTION ACTION REQUIRED
1 Inverter Fault. Reset Unit (1), Call for Service
2 No or low mA signal during exposure Call for Service
3 Exposure terminated by operator Reset unit
4 Stuck or shorted exposure switch. Check or replace switch.
5 Reserved
6 Reserved
7 Reserved
8 Exposure watchdog failure Call for Service
9 Flash CRC Error Call for Service
10 RAM Memory Error Call for Service
11 EEPROM CRC Error Call for Service
12 EEPROM Write Error Call for Service
13 Serial Communications Error Reset unit
14 Stuck or Shorted Prep Switch Check or replace switch.