EasyManua.ls Logo

Sri PXS-710D - Page 67

Sri PXS-710D
73 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
PXS-710D Section 5
65
4.3
CONTINUED:
If an Error Occurs or the User releases the PREP and/or XRAY switches the
Exposure terminates. Otherwise the Exposure completes normally with the
Techniques that were set.
Once the Exposure is completed, a POSTGENCMD is sent to the Host. The
POSTGENCMD contains the Techniques used for the exposure including Time. If
the exposure terminated prematurely, the POSTGENCMD reports the actual
exposure time and returns the Error that caused the termination.
After the Exposure and when the User has released the PREP and XRAY
switches, the Generator returns to the Generator Idle State.
Figure 4.3

Table of Contents