PXS-710D Section 1
17
1.8 CONTINUED:
NOMINAL ELECTRIC POWER
700 W, Constant potential derived from a high frequency inverter source,
Regulated for both kVp and mA.
MAX OUTPUT POWER
Instantaneous: 700W (70kV @ 10 mA)
DUTY CYCLE (AUTOMATIC)
3.3% (ie.. X-ray Exposure 1 second ON, 30 seconds OFF)
MINIMUM PERMANENT FILTRATION
1.8 mm of aluminum equivalent at 70 kVp
REFERENCE CURRENT TIME PRODUCT (Power On Defaults)
Beam Current: 10 mA
Tube Potential: 60 kVp
mAs Setting: 1.60 mAs
LEAKAGE TECHNIQUE FACTORS
Testing is performed at 70 kVp and 10 mA, 10 mAs exposure.
COOLING CURVES
Automatic inhibit circuits provide a delay between exposures proportional to the
exposure time. This delay is sufficient to allow proper tube cooling. The delay
between exposures varies between 2 seconds and 60 seconds depending on the
exposure time.