T15h Tag Deployment & User Guide
71
Sensor Summary Report
This report shows a summary of sensor data (current, minimum, maximum, and
average temperature & humidity values) received by sensor tags during a
selected period.
The report also includes MKT (Mean Kinetic Temperature) values which are
widely used in the pharmaceutical industry. Mean Kinetic Temperature is
calculated for understanding the effects of variations in temperature for
temperature-sensitive goods during storage and transportation. MKT for a
substance can be defined as the calculated value derived from the complete
recording over a specified period of time. If this value is within the permitted
min/max values, the substance is still considered to be safe.
The Sensor Summary Report can show up to 5 years of data within a 3 month
period. Values that exceed thresholds are shown in Red. A Red average value
indicates a persisting problem.
The parameter Only units reporting out of range, allows you to limit the report
to units that reported temperatures or humidity levels below or above the
acceptable range. If you do not check the box, the report will include all the
units that meet the selection criteria.