Error
Code
Error Message
Description
41 Sonic generator low
power
Switch the equipment off, then on again. If problem
persists, contact supplier to arrange a qualified technician
inspection.
42 Maximum cycle time
expired
Make sure hot and cold water flow rates are sufficient.
Make sure filter is not clogged.
Make sure instruments are not blocking drain.
Repeat cycle.
If problem persists, contact supplier to arrange a qualified
technician inspection.
43 Sonic generator serial
comms lost
Switch the equipment off, then on again.
If problem persists, contact supplier to arrange a qualified
technician inspection.
44 Sonic cleaning water
level low
Switch the equipment off, then on again. If problem
persists, contact supplier to arrange a qualified technician
inspection.
45 Lost remote control
with sonic module(s)
Switch the equipment off, then on again.
If problem persists, contact supplier to arrange a qualified
technician inspection.
46 Sonic generator
parameters invalid
Switch the equipment off, then on again.
If problem persists, contact supplier to arrange a qualified
technician inspection.
47 Sonic generator
module PCB fault
Switch the equipment off, then on again.
If problem persists, contact supplier to arrange a qualified
technician inspection.
49 Micro-controller error
Switch the equipment off, then on again. If problem
persists, contact supplier to arrange a qualified technician
inspection.
50 Internal ambient
temperature too high
Make sure room temperature is below +40 °C (+104 °F).
Make sure cooling fan air intake is not blocked or
restricted. Turn equipment off and leave for 1 hour to cool
down.
If problem persists, contact supplier to arrange a qualified
technician inspection.
Table 11.1 Error Codes and Remedy Actions (continued)
TROUBLESHOOTING 93
InnoWave Unity & Unity 20 Operator Manual