Description
51 Fill level sensor invalid
Repeat cycle.
If problem persists, contact supplier to arrange a qualified
technician inspection.
52 Real-time clock read
error
Switch the equipment off, then on again.
If problem persists, contact supplier to arrange a qualified
technician inspection.
53 24V power supply
voltage out-of-range
Switch the equipment off, then on again.
If problem persists, contact supplier to arrange a qualified
technician inspection.
57 Micro-controller
stack overflow
Repeat cycle. If problem persists, contact supplier to
arrange a qualified technician inspection.
58 EEPROM read error
Repeat cycle. If problem persists, contact supplier to
arrange a qualified technician inspection.
59 Unable to store cycle
log
Repeat cycle. If problem persists, contact supplier to
arrange a qualified technician inspection.
Table 11.1 Error Codes and Remedy Actions (continued)
94 TROUBLESHOOTING
InnoWave Unity & Unity 20 Operator Manual