EasyManua.ls Logo

Steris InnoWave Unity - 12.3 Soak Cycle Program

Steris InnoWave Unity
146 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Error
Code
Error Message
Description
51 Fill level sensor invalid
Repeat cycle.
If problem persists, contact supplier to arrange a qualified
technician inspection.
52 Real-time clock read
error
Switch the equipment off, then on again.
If problem persists, contact supplier to arrange a qualified
technician inspection.
53 24V power supply
voltage out-of-range
Switch the equipment off, then on again.
If problem persists, contact supplier to arrange a qualified
technician inspection.
57 Micro-controller
stack overflow
Repeat cycle. If problem persists, contact supplier to
arrange a qualified technician inspection.
58 EEPROM read error
Repeat cycle. If problem persists, contact supplier to
arrange a qualified technician inspection.
59 Unable to store cycle
log
Repeat cycle. If problem persists, contact supplier to
arrange a qualified technician inspection.
Table 11.1 Error Codes and Remedy Actions (continued)
94 TROUBLESHOOTING
InnoWave Unity & Unity 20 Operator Manual

Table of Contents

Related product manuals